1: \begin{abstract}
2: The use of Time Of Flight (TOF) in Positron Emission Tomography (PET)
3: is expected to reduce noise on images, thanks to the additional information.
4: In clinical routine, a common reconstruction approach is the use of
5: maximum likelihood expectation maximization (MLEM) stopped after few
6: iterations. Empirically it was reported that, at matched number of
7: iterations, the introduction of TOF increases noise. In this work
8: we revise the theory describing the signal and noise convergence in
9: MLEM, and we adapt it to describe the TOF impact on early stopped
10: MLEM. We validated theoretical results using both computer simulations
11: and phantom measurements, performed on scanners with different coincidence
12: timing resolutions. This work provides theoretical support for the
13: empirically observed noise increase introduced by TOF. Conversely,
14: it shows that TOF not only improves signal convergence but also makes
15: it less dependent on the activity distribution in the field of view.
16: We then propose a strategy to determine stopping criteria for TOF-MLEM,
17: which reduces the number of iterations by a factor proportional to
18: the coincidence timing resolution. We prove that this criteria succeeds
19: in markedly reducing noise, while improving signal recovery robustness
20: as it provides a level of contrast recovery which is independent from
21: the object dimension and from the activity distribution of the background.