1: % Template article for preprint document class `elsart'
2: % SP 2001/01/05
3:
4: \documentclass{elsart}
5:
6: % Use the option doublespacing or reviewcopy to obtain double line spacing
7: % \documentclass[doublespacing]{elsart}
8:
9: % if you use PostScript figures in your article
10: % use the graphics package for simple commands
11: % \usepackage{graphics}
12: % or use the graphicx package for more complicated commands
13: % \usepackage{graphicx}
14: % or use the epsfig package if you prefer to use the old commands
15: \usepackage{rotating}
16: \usepackage{epsfig}
17: \usepackage{cite}
18:
19: % The amssymb package provides various useful mathematical symbols
20: \usepackage{amssymb}
21:
22: % \parskip12pt plus 1pt minus 1pt
23: % \parindent 0mm
24: % \setlength{\floatsep}{0em}
25:
26: \setcounter{bottomnumber}{4}
27: \setcounter{totalnumber}{9}
28: \renewcommand{\bottomfraction}{1.0}
29: \renewcommand{\textfraction}{0.0}
30: \renewcommand{\arraystretch}{1.1}
31:
32: \def\mum{$\mu$m}
33: \def\be{\begin{equation} }
34: \def\ee{\end{equation} }
35: \def\ba{\begin{eqnarray} }
36: \def\ea{\end{eqnarray} }
37: \def\ban{\begin{eqnarray*} }
38: \def\ean{\end{eqnarray*} }
39: \def\epem{\mbox{e}^+\mbox{e}^-}
40: \def\arco{\mbox{ArCO$_2$}}
41: \def\mmrcm{\mbox{$\mu$m / $\sqrt{\mbox{cm}}$}}
42: \begin{document}
43:
44: \begin{frontmatter}
45:
46: % Title, authors and addresses
47:
48: % use the thanksref command within \title, \author or \address for footnotes;
49: % use the corauthref command within \author for corresponding author footnotes;
50: % use the ead command for the email address,
51: % and the form \ead[url] for the home page:
52: % \title{Title\thanksref{label1}}
53: % \thanks[label1]{}
54: % \author{Name\corauthref{cor1}\thanksref{label2}}
55: % \ead{email address}
56: % \ead[url]{home page}
57: % \thanks[label2]{}
58: % \corauth[cor1]{}
59: % \address{Address\thanksref{label3}}
60: % \thanks[label3]{}
61: \title{Position Sensing from Charge Dispersion in
62: Micro-Pattern Gas Detectors with a Resistive Anode}
63:
64: % \author[ref_CU]{B.~Carnegie},
65: \author[ref_CU,ref_TR]{M.~S.~Dixit\corauthref{add_MSD}},
66: \ead{msd@physics.carleton.ca}
67: \author[ref_JD]{J.~Dubeau},
68: \author[ref_UM]{J.-P.~Martin}
69: % \author[ref_CU]{H.~Mes},
70: and
71: \author[ref_CU]{K.~Sachs}
72: % \ead{sachs@physics.carleton.ca}
73: % \ead[url]{home page}
74:
75: \address[ref_CU]{Department of Physics, Carleton University,
76: \\ 1125 Colonel By Drive, Ottawa, ON K1S 5B6 Canada}
77: \address[ref_JD]{DETEC, Aylmer, QC, Canada}
78: \address[ref_UM]{Universiy of Montreal, Montreal, QC, Canada}
79: \address[ref_TR]{TRIUMF, Vancouver, BC Canada}
80: \corauth[add_MSD]{Corresponding author.
81: Tel.: +1-613-520-2600, ext. 7535; fax: +1-613-520-7546.}
82: % \corauth[add_KS]{Corresponding author;
83: % tel.: +1-613-520-2600, ext. 1567; fax: +1-613-520-7546.}
84:
85:
86: % use optional labels to link authors explicitly to addresses:
87: % \author[label1,label2]{}
88: % \address[label1]{}
89: % \address[label2]{}
90:
91: \begin{abstract}
92: % Text of abstract
93: Micro-pattern gas detectors, such as the Gas Electron Multiplier (GEM) and the Micromegas
94: need narrow high density anode readout elements to achieve good spatial resolution. A
95: high-density anode readout would require an unmanageable number of electronics channels for certain
96: potential micro-detector applications such as the Time Projection Chamber. We describe below a
97: new technique to achieve good spatial resolution without increasing the electronics channel
98: count in a modified micro-detector outfitted with a high surface resistivity anode readout
99: structure. The concept and preliminary measurements of spatial resolution from charge
100: dispersion in a modified GEM detector with a resistive anode are described below.
101: \end{abstract}
102:
103: \begin{keyword}
104: % keywords here, in the form: keyword \sep keyword
105: Gaseous Detectors \sep
106: Position-Sensitive Detectors \sep
107: Micro-Pattern Gas Detectors \sep
108: Gas Electron Multiplier \sep
109: Micromegas
110:
111: % PACS codes here, in the form: \PACS code \sep code
112: \PACS 29.40.Cs \sep 29.40.Gx
113:
114: \end{keyword}
115: \end{frontmatter}
116:
117: % main text
118: \section{Introduction}
119: \label{sec:intro}
120:
121: A new class of high-resolution multi-channel gas avalanche micro-detectors has been
122: developed during the past decade for charged particle tracking. The Gas Electron Multiplier
123: (GEM) \cite{cit:GEM} and the Micromegas \cite{cit:MM} are examples of some of the new micro-pattern gas
124: detectors \cite{cit:mpgd} already in wide use. The GEM and the Micromegas sample the avalanche charge
125: using arrays of closely spaced long parallel anode strips to measure a single co-ordinate.
126: Spatial resolutions of 40 to 50 \mum\ are typical with anodes strips at 200 \mum\ pitch.
127: Micro-detectors do require more instrumented channels of electronics than multi-wire proportional
128: chambers. However, the number of readout channels has not yet become an issue for most
129: experiments using micro-detectors for charged particle tracking.
130:
131: There are potential micro-detector applications, however, where the electronics channel count
132: may become an issue. The Time Projection Chamber (TPC) \cite{cit:TPC} used in high energy physics
133: experiments is one such example. A single endcap detector is used in the TPC to measure
134: both the radial and the azimuthal co-ordinates of the ionization charge cluster. In the
135: conventional TPC, read out with a multi-wire proportional chamber endcap, the radial
136: co-ordinate is obtained from the anode wire position. A precise second co-ordinate along the
137: anode wire length is measured by sampling the induced cathode charge with a series of
138: several mm wide rectangular pads.
139:
140: The spatial resolution of a TPC in a high magnetic field is dominated by the
141: {\boldmath $E\times B$\unboldmath} and track
142: angle systematic effects \cite{cit:clif,cit:amend}. Replacement of the usual anode wire/cathode pad readout with
143: one based on micro-detectors with anode pad readout would almost entirely eliminate the
144: systematics and has the potential to improve the TPC performance significantly. However, the
145: suppression of the transverse diffusion in a high magnetic field may often result in the
146: collection of most of the avalanche charge within the width of a single anode pad resulting in
147: a loss of TPC resolution. For better resolution, a micro-detector readout TPC will need either
148: a finely segmented anode pad structure with a prohibitively large number of instrumented
149: channels of electronics, or perhaps the complication of specially shaped pads which enhance
150: anode charge sharing \cite{cit:chevron}.
151:
152: We describe here a new technique which can be used to measure the position of a localized
153: charge cluster in a micro-detector using pads of widths similar to those used in wire-pad
154: systems. Most of our tests have so far been done with a modified double-GEM detector.
155: However, the new technique appears to be sufficiently general to be applicable to the
156: Micromegas.
157:
158:
159: \section{Charge dispersion in a micro-detector with a resistive anode}
160:
161: With certain modifications to the anode readout structure, it is possible to measure the position
162: of a localized charge cluster in a micro-detector with pads wider than have been used so far. A
163: thin high surface resistivity film is glued to a separate readout pad plane and is used for the
164: anode (Fig. \ref{fig:scheme}). The resistive anode film forms a distributed 2-dimensional resistive-capacitive
165: network with respect to the readout pad plane. Any localized charge arriving at the anode surface
166: will be dispersed with the $RC$ time constant determined by the anode surface resistivity and the
167: capacitance density determined by the spacing between the anode and readout planes. With the
168: initial charge dispersed and covering a larger area with time, wider pads can be used for signal
169: pickup and position determination.
170:
171: This spatial dispersion, which can be explained by a simple physical
172: model described below, does not attenuate the pad charge signal.
173: As the initial charge cluster diffuses, the electron density on the
174: resistive surface, varying with time, is capacitively mirrored on the readout
175: pads below. Since the readout pads are directly connected to the preamplifiers,
176: there is no charge signal amplitude loss.
177:
178: \subsection{A model for charge dispersion in a micro-detector with a resistive anode}
179:
180: The resistive anode and the readout plane together can be looked upon to form a distributed
181: 2 dimensional $RC$ network in the finite element approximation. Consider first the 1 dimensional
182: problem of a point charge arriving at $t=0$ at the origin in the middle of an infinitely long wire
183: grounded at both ends. For small inductance, the space-time evolution of the charge density
184: $\rho$ on the wire is given by the well-known Telegraph equation:
185: \be
186: \frac{\partial \rho}{\partial t} = h \frac{\partial^2 \rho}{\partial x^2}
187: \hspace{1cm}\mbox{where} \hspace{1cm}
188: h = 1/RC \; . \label{eq:T-1dim} \ee
189: Here $R$ is resistance per unit length and $C$ the capacitance per unit length for the wire.
190:
191: The solution for charge density is given by:
192: \be
193: \rho(x,t) = \sqrt{\frac{1}{4\pi t h}} \exp(-x^2 / 4 t h)\; . \label{eq:1dim} \ee
194:
195:
196: In analogy with the 1 dimensional case, we can write the Telegraph equation for the case of a
197: resistive surface. At time $t = 0$, a point charge is collected at the origin by a resistive anode
198: surface of infinite radius (for simplicity). The 2-dimensional Telegraph equation for the charge
199: density is:
200: \be
201: \frac{\partial \rho}{\partial t} = h \left[ \frac{\partial^2\rho}{\partial r^2} +
202: \frac{1}{r}\frac{\partial \rho}{\partial r} \right] \; , \label{eq:T-2dim} \ee
203: where in this case, $R$ is the surface resistivity and $C$ is capacitance per unit area.
204:
205: The solution for the charge density function in this case is given by:
206: \be
207: \rho(r,t) = \frac{1}{2 t h} \exp(-r^2 / 4 t h)\; . \label{eq:2dim} \ee
208:
209:
210: The charge density function (equation \ref{eq:2dim}) for the resistive anode varies with time and is
211: capacitively sampled by the readout pads. Fig. \ref{fig:charge} shows the time evolution of the charge density
212: for an initially localized charge cluster for our detector. The charge
213: signal on a pad can be computed by integrating the time dependent charge density function over
214: the pad area. The shape of the charge pulse on a pad depends on the pad geometry, the location
215: of the pad with respect to the initial charge and the $RC$ time constant of the system.
216:
217:
218: \subsection{Charge dispersion signal in micro-detectors with long readout strips}
219:
220: The charge dispersion measurements were carried out with a modified GEM detector with long
221: anode strips. Since a spatial co-ordinate measurement for long strips is meaningful only in a
222: direction transverse to the strip length, we can use the 1-dimensional Telegraph equation
223: (\ref{eq:T-1dim}) to
224: describe the situation. However, the solution for the charge density must account for the finite
225: size of the resistive anode in contrast to the solution given by equation (\ref{eq:1dim}) in the long wire
226: approximation.
227:
228: The boundary conditions to solve equation (\ref{eq:T-1dim}) in this case are:
229: \be
230: \rho(x=0,t) = \rho(x=s,t) = 0\; ; \; 0 \leq t \leq \infty \; , \label{eq:boundary} \ee
231: where $s$ is the size of the resistive foil (assumed square) held at ground potential along its
232: boundary.
233:
234: The solution satisfying the finite boundary conditions is:
235: \be
236: \rho(x,t) = \sum_{m=1}^\infty A_m \exp [ -h(m\pi/ s)^2 t] \sin(xm\pi/s) \; , \label{eq:simsol} \ee
237: where the coefficients $A_m$ can be determined from the knowledge of the initial charge density:
238: \be
239: A_m = \frac{2}{s}\int_0^s \rho(x,t=0) \sin(xm\pi/s) dx \; . \label{eq:coeff} \ee
240:
241: The signal on a readout strip can be computed by integrating the charge density function over
242: the strip width. Furthermore, the finite extent of the initial charge cluster, the intrinsic
243: micro-detector charge signal rise time as well as the rise and fall time characteristics of the front-end
244: electronics determine the shape of the measured signal shape. All these parameters need to be
245: included in the model to compare to the experiment.
246:
247: Model calculations were done for a GEM detector with a resistive anode readout with 1 mm
248: wide strips. The anode resistivity and anode-readout gap limits
249: the computed spatial spread of the charge dispersion over pads to about 700~\mum\ comparable to
250: transverse diffusion in a high magnetic field TPC. Simulated signals for the readout strip directly
251: below the initial ionization charge cluster and for the next four adjacent strips are shown in
252: Fig. \ref{fig:simsig}. The same figure also shows the simulated pad response function or equivalently the
253: spatial spread of an initially localized charge cluster due to charge dispersion.
254:
255:
256: \section{Spatial resolution measurements in a GEM detector with a resistive anode}
257:
258: The charge dispersion test measurements were made with the modified double-GEM detector
259: (Fig. \ref{fig:scheme}) filled with Ar/CO$_2$ 90/10. A 50 \mum\ thick Mylar film with a
260: surface resistivity of 2.5 M$\Omega$
261: per square was glued to the pad readout board. The spacing between the anode and readout
262: planes was close to 100~\mum\ including the thickness of the glue.
263:
264: The initial ionization is provided by x-ray photon conversion in the gas.
265: The average photon energy was
266: about 4.5 keV as low energy bremsstrahlung photons from the copper target x-ray tube, run at 7
267: kV, were absorbed by the material in the x-ray tube and detector windows. A
268: \mbox{$\sim$40 \mum} pinhole in
269: a thin brass sheet was used to produce a miniaturized x-ray tube focal spot image in the GEM
270: drift gap. The size of the x-ray spot at the detector is estimated to be on the order of
271: 70 \mum .
272: After avalanche multiplication and diffusion, the RMS size of the electron cloud reaching the
273: resistive anode was \mbox{$\sim$400 \mum}. The gas gain was about 2500.
274:
275: Signals were read out from 7 cm long and 1.5 mm wide strips. The front-end electronics
276: consisted of Aleph \cite{cit:amend} TPC wire charge preamplifiers followed by receiver amplifiers. Signals
277: from 8 contiguous strips were digitized using two 4-channel Tektronix digitizing oscilloscopes.
278: A computerized translation stage was used to move the x-ray spot in small steps over the width
279: of the centre strip. One thousand event runs were recorded for each x-ray spot position on an
280: event by event basis.
281:
282: For a given anode surface resistivity and readout geometry, the observed shape of the charge
283: pulse depends on the strip position with respect to the location of primary charge cluster on the
284: resistive anode. Fig. \ref{fig:scope} shows an event where the x-ray ionization spot is located directly above
285: the centre of a readout strip. A fast charge pulse is observed on the strip peaking in time with the
286: maximum of the charge density at the anode surface above. Pulses on strips farther away have a
287: slower rise time and peak late because the local charge density on the anode surface nearest the
288: strip reaches its maximum later. Also, an early short duration induced pulse is visible for strips
289: adjacent to the main strip. The induced pulses \cite{cit:orsay}, produced by electron motion in the GEM
290: induction gap, have demonstrated position sensitivity \cite{cit:LCWS} but require the use of high-speed pulse
291: shape sampling electronics. In addition, measurable induced pulses are specific to GEM
292: detectors with sizeable induction gaps. For charge dispersion measurements described below, the
293: induced GEM pulse information has not been used.
294:
295:
296:
297: \section{Data analysis and results}
298:
299: The charge dispersion signals were confined to a narrow region on the readout board in the
300: present setup. There were measurable signals above noise only on three 1.5 mm wide strips. The
301: analysis to determine the space point resolution from the event by event data consisted of
302: following steps: a) Determine the pulse heights of signals on the strips, independent of rise time;
303: b) Compute a centre of gravity position for the event from the measured pulse heights; c) Correct
304: for the bias in the computed centre of gravity to obtain the position of the ionization cluster for
305: the event. The standard deviation of the bias corrected centre of gravity position with respect to the
306: known x-ray beam spot position for the run gives a measure of the space point resolution.
307:
308: The data from each 1000 event run were sub-divided into two equal data sets: one used for
309: calibration and one for resolution studies. The pulse heights were obtained by fitting polynomial
310: functions to the digitized pulse shape data. The calibration data set was used to determine and fix
311: the coefficients of polynomial functions (see Fig. \ref{fig:poly}) used subsequently in the analysis of events
312: in the resolution data set.
313:
314: The centres of gravity for the events in the calibration data set were computed from the measured
315: pulse heights. The correction function for the bias in the centre of gravity method
316: (see Fig. \ref{fig:bias})
317: was determined by plotting the mean value of computed centres of gravity against the known
318: x-ray spot position for the individual runs.
319:
320: The pulse heights and the peak positions for the events used for resolution studies were
321: determined by fitting the pulses to the fixed polynomials shapes obtained from the corresponding
322: calibration data set. The computed centres of gravity were converted to ``true position'' by
323: interpolation using the bias correction function determined from the calibration data set.
324: Fig. \ref{fig:resolution}
325: shows the measured resolution function for the 1.5 mm wide strips at two different positions of
326: the x-ray ionization spot over the strip. Fig. \ref{fig:summary} shows the measured spatial resolutions and the
327: position residuals: i.e. the deviation of measured positions from the micrometer settings. The
328: small systematic trends apparent in the plot appear to be related to an imprecise knowledge of the
329: experimental parameters in the present setup and any remaining biases in the present method of
330: analysis. Nevertheless, the standard deviations of the position measurements, in the range of 50
331: to 80 \mum , are all consistent with the size of the collimated x-ray spot at the detector.
332:
333:
334: \section{Outlook and summary}
335:
336: We have demonstrated that a controlled $RC$ dispersion of the avalanche charge makes it possible
337: to measure its position with a micro-detector with strips wider than have been used previously.
338: The pad response function and signal shapes are determined by the anode surface resistivity and
339: anode-readout plane gap. With the proper choice of the $RC$ time constant of the system, the
340: charge dispersion technique will not compromise the counting rate ability of the detector. Nor
341: should it compromise the 2-track resolving power of the detector which should be limited only
342: by the diffusion effects in the gas. Once the characteristics of the charge dispersion signal are
343: properly understood, it should also be possible to simplify the technique replacing the pulse
344: shape measurement system with less expensive charge integrating electronics.
345:
346: The charge dispersion space point resolution studies described here were done for long readout
347: strips in a modified GEM detector with a resistive anode. Further experimental and simulation
348: studies are in progress to investigate the spatial resolution capabilities of the charge dispersion
349: technique with rectangular pads similar to those used in wire-cathode pad TPCs. We are also
350: testing the concept of position sensing from charge dispersion with the Miromegas where the
351: high anode resistance may help improve the detector HV stability as well as protect the front-end
352: electronics from spark damage.
353:
354: \clearpage
355:
356: \section*{Acknowledgments}
357:
358: The charge preamplifiers used in these measurements came from the Aleph TPC at CERN and
359: we wish to thank Ron Settles for making these available to us. Ernie Neuheimer lent us his
360: expertise in designing, building and troubleshooting much of the specialized electronics used for
361: these measurements. Mechanical engineers Morley O'Neill and Vance Strickland helped with
362: the detector design and in improving the clean-room facility used for the detector assembly.
363: Philippe Gravelle provided technical assistance when needed. Our CO-OP students Alasdair
364: Rankin and Steven Kennedy made significant contributions to all aspects of this research from
365: hardware construction to improving the data acquisition software as well as writing some of the
366: data analysis code. Finally, one of the authors (MSD) would like to express his thanks to
367: V.~Radeka for an illuminating discussion concerning the phenomenon of charge dispersion. This
368: research was supported by a project grant from the Natural Sciences and Engineering Research
369: Council of Canada.
370:
371: \begin{thebibliography}{00}
372: \bibitem{cit:GEM} F. Sauli, Nucl. Inst. Meth. {\bf A386} (1997) 531.
373: \bibitem{cit:MM} Y. Giomataris et al, Nucl. Inst. Meth. {\bf A376} (1996) 29.
374: \bibitem{cit:mpgd} F. Sauli and A. Sharma, Ann. Rev. Nucl. Part. Sci. {\bf 49} (1999) 341.
375: \bibitem{cit:TPC} D. R. Nygren, PEP 198 (1975).
376: \bibitem{cit:clif} C.K.Hargrove et al, Nucl. Inst. Meth. {\bf 219} (1984) 461.
377: \bibitem{cit:amend} S. R. Amendolia et al., Nucl. Inst. Meth. {\bf A283} (1989) 573.
378: \bibitem{cit:chevron} M. Schumacher, LC-DET-2001-014.
379: \bibitem{cit:orsay} M.S. Dixit et al, Proceedings of Workshop on Micro-Pattern Gas Detectors, Orsay France
380: (1999).
381: \bibitem{cit:LCWS} D. Karlen et al, Physics and experiments with future linear $\rm e^+e^-$ colliders, LCWS2000,
382: American Institute of Physics Conf. Proc. Vol 578.
383:
384: % \bibitem{label}
385: % Text of bibliographic item
386: % notes:
387: % \bibitem{label} \note
388: % subbibitems:
389: % \begin{subbibitems}{label}
390: % \bibitem{label1}
391: % \bibitem{label2}
392: % If there is a note, it should come last:
393: % \bibitem{label3} \note
394: % \end{subbibitems}
395: % \bibitem{}
396: \end{thebibliography}
397:
398: \clearpage
399:
400:
401:
402:
403: \begin{figure}[bp]
404: \centerline{\mbox{\epsfxsize=\textwidth \epsffile{fig1_GEM_foil.eps}}}
405: \caption[]{\label{fig:scheme}
406: Schematics of the resistive anode double-GEM detector used for charge dispersion
407: studies.
408: }
409: \end{figure}
410:
411: \begin{figure}[bp]
412: \centerline{\mbox{\epsfxsize=\textwidth \epsffile{fig2_dispersion.eps}}}
413: \caption[]{\label{fig:charge}
414: The evolution of the two dimensional charge density function (equation \ref{eq:2dim})
415: on the resistive anode with a resistivity of 2.5~M$\Omega$ per
416: square and anode-readout plane separation of 100~\mum .
417: For the model calculation the initial
418: charge was point-like and localized at the origin at time = 0.
419: }
420: \end{figure}
421:
422: \begin{figure}[bp]
423: \centerline{\mbox{\epsfxsize=\textwidth \epsffile{fig3_padres.eps}}}
424: \caption[]{\label{fig:simsig}
425: Simulated signals for 1 mm wide strips for an anode resistivity of
426: 2.5~M$\Omega$ per
427: square and anode-readout plane separation of 100~\mum . The pad response function for charge
428: dispersion is shown on the right. The diffusion effects were neglected for this simulation.
429: }
430: \end{figure}
431:
432:
433:
434: \begin{figure}[bp]
435: \epsfxsize=\textwidth \epsffile{fig5_ScopePulses.eps}
436: \caption[]{\label{fig:scope}
437: Observed charge dispersion signals on three adjacent strips for a single x-ray photon conversion in the
438: double-GEM detector. Tektronix scope pulses with \mbox{400 ns/div} on the
439: a)~right strip (20 mV/div),
440: b)~central strip (50 mV/div) and
441: c)~left strip (20 mV/div).}
442: \end{figure}
443:
444: \begin{figure}[bp]
445: \centerline{\mbox{\epsfxsize=\textwidth \epsffile{fig6_scope.eps}}}
446: \caption[]{\label{fig:poly}
447: Polynomials were fitted to the calibration data set for each run to fix the functions used to
448: determine the pulse height of events for resolution studies. The top figure shows the polynomial
449: fit to the average of fast charge pulses for the centre strip. The figure below shows the
450: polynomial fit to the average of late charge dispersion pulses for an adjacent strip.
451: }
452: \end{figure}
453:
454:
455:
456:
457: \begin{figure}[bp]
458: \centerline{\mbox{\epsfxsize=10cm \epsffile{fig7_bias.eps}}}
459: \caption[]{\label{fig:bias}
460: The bias correction function shown in the figure was experimentally determined. The bias
461: function was used in converting the computed centroid of signals on three strips to the true
462: position on an event by event basis.
463: }
464: \end{figure}
465:
466:
467:
468:
469: \begin{figure}[bp]
470: \centerline{\mbox{\epsfxsize=\textwidth \epsffile{fig8_centroid.eps}}}
471: \caption[]{\label{fig:resolution}
472: The resolution function for two x-ray beam spot positions.
473: }
474: \end{figure}
475:
476:
477:
478: \begin{figure}[bp]
479: \centerline{\mbox{\epsfxsize=\textwidth \epsffile{fig9_resolution.eps}}}
480: \caption[]{\label{fig:summary}
481: The summary of the spatial resolutions and the position residuals for the x-ray scan across
482: the readout strips.
483: }
484: \end{figure}
485:
486: \end{document}
487:
488: